Fluke

Fluke Fluke 985 Fluke 985 Particle Counter Fluke 985

  • MPN: Fluke 985
  • UPC: 095969620321
  • PPN: FLU-FLUKE-985

Quote Required

Available to Order from Manufacturer

Lead times may vary. For an accurate estimate please contact us.

Available To Ship Immediately:0

Looking to buy a large quantity?

To View Branch Inventory And Your Custom Pricing

Spec sheet

Description

Fluke 985 Particle Counter Fluke 985

Features

  • Assures accurate measurements even for critical locations requiring ISO Class 5 – 9 certification
  • Promises easy single-hand operation in tight or awkward spaces with less operator fatigue
  • Ten hours of standard-use battery life to last a full work day
  • Easy navigation and viewing with the intuitive icons and large font options
  • Quick access to historical data
  • Data can be viewed in a traditional tabular or trend graph view at the user’s preference
  • Customizable settings for the display, sample methods, and alarm settings for sample count
  • Download data to a PC using a USB memory stick, USB cable or Ethernet connection. No special software is required to download or view data
  • Confirms security when required
  • Assures device is always ready to go when needed
  • Assess air quality more easily and accurately.

Documentation & Downloads

Specifications

Measured Specifications

Channels0.3 μm, 0.5 μm, 1.0 μm, 2.0 μm, 5.0 μm, 10.0 μm
Light source775 nm to 795 nm, 90 mW class 3B laser
Counting efficiency50% @ 0.3 μm; 100% for particles > 0.45 μm (per ISO 21501)
Concentration limits10% at 4,000,000 particles per ft3(per ISO 21501)
AlarmsCounts, low battery, sensor fail
Sample inletIsokinetic probe
Communication modeUSB or Etheret
Weight0.68 kg (1.5 lb)
Rechargeable batteryLi-ion 7.4 V 2600 mAh; 10 hours typical use (5 hours continuous sampling) / 3.5 hours
Operating environment50°F to 104°F (10°C to 40°C) / < 95% non-condensing relative humidity
WarrantyOne-year
Flow rate0.1 cfm (2.83 L/min)
Count modesRaw counts, #/m3, #/ft3, #/liter in Cumulative or Differential mode
Zero count1 count / 5 minute (JIS B9921)
Data storage10,000 records (rotating buffer)
Delay time0 to 24 hours
SecurityAdministrator password control (optional)
Dimensions (L x W x D)27.2 x 9.9 x 5.3 cm (10.7 x 3.9 x 2.1 in)
Power adapter100 V AC to 240 V AC, 12 V DC, 2.5 A
CalibrationPSL particles in air (NIST traceable)
Storage environment14°F to 122°F (-10°C to 50°C)/Up to 98% non-condensing relative humidity
Accessories includedCradle for charging and convenient USB and Ethernet communications, USB-A to MINI-B 6 foot cable, 12 V dc power supply, zero count inlet-filter, filter adapter, sample inlet protective cap, hard case
Fluke 985 Particle Counter Fluke 985
banner

Need assistance?

We can help! Connect with our Sales Team Today!