561 HVAC Infrared & Contact Thermometer Fluke 561
Fluke

Fluke FLUKE 561 561 HVAC Infrared & Contact Thermometer Fluke 561

  • MPN: FLUKE 561
  • UPC: 095969332538
  • PPN: FLU-FLUKE- 561

CAD

$437.17

/ EA

Available to Order

Lead times may vary. For an accurate estimate please contact us.

Available To Ship Immediately:0

Looking to buy a large quantity?

To View Branch Inventory And Your Custom Pricing

Spec sheet

Description

561 HVAC Infrared & Contact Thermometer Fluke 561

Features

  • Contact and non-contact IR thermometer for measuring up close or at a distance, without a ladder
  • Measures temperature from -40°C to 550° C (-40°F to 1022°F)
  • Works with all standard mini-connector type K thermocouples
  • Provides single-point laser sighting
  • Allows easy emissivity adjustment for measuring pipes and other shiny surfaces more accurately

Documentation & Downloads

Specifications

General Specifications

Temperature
-40 to 550°C (-40 to 1022°F)
D:S (distance to spot size)
12:1
Display accuracy [assumes ambient operating temperature of 23°C (73°F) to 25°C (77°F)]
±1.0% of reading ±1% of reading or ±1°C (±2°F), whichever is greater below 0°C / 32°F, ±1°C (±2°F) ±0.1°/1°
Repeatability
±0.5% of reading or ±1°C (±2°F), whichever is greater
Laser sighting
Single point laser
Laser power
Class 2(II) operation; Output < 1 mW, wavelength 630-670 nm
Power, battery life
2 AA batteries (alkaline or NiCD)
Display hold
7 seconds
Warranty
2 years
Storage temperature
-20°C to 65°C (-25°F to 150°F)
Thermocouple Type K mini adapter input
Yes, compatible with industry standard K type probes with mini-connector. Displays to the thermometer temperature range.
HVAC measurement guide
Yes
Display resolution0.1°C (0.1°F) of reading
Easy emissivity selectorLow (0.3), medium (0.7). high (0.95)
Response time500 mSec (95% of reading)
Spectral response8 µm to 14 µm
Laser shutoffLaser turns off above ambient temperature of 40°C (104°F)
Relative humidity10% to 90% RH non-condensing, at < 30°C (86°F)
Battery life12 hours
Backlit displayYes, LCD with dual temperatures (current and MAX/MIN/DIF/KTC), low battery, F/C indicator, and Scan/Hold options
Operating temperature0°C to 50°C (32°F to 120°F)
MAX, MIN, DIF temperaturesYes
Thermocouple Type K fabric hook-and-loop fastener pipe probeYes, with a temperature range of 0°C to 100°C (32°F to 212°F) and accuracy of ±2.2°C (±4°F)

Similar Products

Loading carousel layout...

CAD$437.17/ EA

Loading add to cart quantity...

banner

Need assistance?

We can help! Connect with our Sales Team Today!